ISO 13485 surveillance: three records to have ready before audit day

T
TAC Editorial Team
•9 November 2026

ISO 13485 surveillance auditors follow real devices through your records. Here are the three records to test first, the medical device file, the design and development file and complaint handling, and how to check them in an afternoon.

An ISO 13485 surveillance audit feels different from an ISO 9001 one, and there is a good reason. ISO 13485 is a quality management standard "for regulatory purposes". Almost every requirement ties back to a specific device, and to the regulations that apply to it.

So the auditor does not just ask whether you have a procedure. They pick a device, or a device family, and ask you to show them its records. If those records are complete, current and easy to find, the audit moves along. If they are spread across three drives and someone's inbox, it does not matter how good the procedure is.

Here are three records at the heart of that sampling, and a simple way to test each one before your certification body does. Clause references are to ISO 13485:2016, still the current edition. Unlike ISO 9001 and ISO 14001, there is no new edition to transition to.

Record 1: the medical device file (clause 4.2.3)

For each medical device type or family, ISO 13485 asks you to establish and maintain one or more files that contain or reference the documents showing conformity to the standard and compliance with applicable regulatory requirements.

The standard lists what the file must include as a minimum:

  • a general description of the device, its intended use or purpose, and its labelling, including any instructions for use

  • product specifications

  • specifications or procedures for manufacturing, packaging, storage, handling and distribution

  • procedures for measuring and monitoring

  • installation requirements, where appropriate

  • servicing procedures, where appropriate.
  • Where it goes wrong. The file was built for initial certification and has not kept up. The label and instructions for use have moved on a revision, packaging has changed supplier, and the file still points at the old versions. Nothing is missing on paper. It is just out of date.

    The check. Take one device off the shelf, or pull its current label and instructions for use from your document system. Now open the medical device file and follow each reference. Does every one lead to the current approved version? Could someone outside your team find each document quickly?

    Record 2: the design and development file, including changes (clauses 7.3.9 and 7.3.10)

    If you design devices, ISO 13485 asks for a design and development file for each device type or family. It must include or reference the records that show conformity to the design and development requirements, and the records of design and development changes.

    The change records are where surveillance auditors tend to dig, because changes keep happening after certification. Clause 7.3.9 asks you to:

  • determine how significant each change is to the function, performance, usability, safety and regulatory requirements of the device and its intended use

  • review, verify, validate (as appropriate) and approve the change before it is implemented

  • evaluate the effect of the change on constituent parts, product in process or already delivered, and the inputs or outputs of risk management and product realisation

  • keep records of the changes, their review and any actions.
  • Where it goes wrong. A small change, often a component swap forced by a supplier, goes through purchasing and production but never through design control. Or it goes through design control, but nobody records why it was judged minor, or whether risk management was looked at.

    The check. Find the last three design changes. For each one, can you show the significance decision, the review, verification and approval dated before the change went live, and a note on risk management? Then look at your purchasing and engineering change records for the same period. Is there any change that touched the device but never reached the design file?

    If design and development is excluded from your system, the same principle applies to the justification. The standard allows the exclusion only where applicable regulatory requirements permit it, and your claims of conformity have to reflect it. Make sure that justification is current and that your certificate scope matches it.

    Record 3: complaint handling and regulatory reporting (clauses 8.2.2 and 8.2.3)

    ISO 13485 asks for documented procedures for timely complaint handling in line with applicable regulatory requirements. At a minimum they must cover:

  • receiving and recording information

  • deciding whether the feedback is a complaint

  • investigating complaints

  • deciding whether the information needs to be reported to the appropriate regulatory authorities

  • handling complaint-related product

  • deciding whether corrections or corrective actions are needed.
  • If a complaint is not investigated, the justification must be documented. Any correction or corrective action from a complaint must be documented. If the investigation finds that an outside party contributed, you exchange information with them. Where regulations require notification of adverse events or advisory notices, clause 8.2.3 asks for a procedure and records of the reporting.

    For devices on the Great Britain market, that reporting sits within the post-market surveillance requirements that came into force on 16 June 2025. The MHRA says serious incidents must now be reported within 15 days, down from 30, with shorter deadlines for the most serious categories set out in its vigilance guidance. Your complaint records need to show the reportability decision was made in time to meet those deadlines.

    Where it goes wrong. The investigation is good, but the reportability decision is not written down, especially when the answer was "not reportable". Or feedback that came in through a sales rep or a distributor email was never logged as a complaint at all.

    The check. Pick three complaints, including one closed as not reportable. For each, can you show when it was received, who decided it was a complaint, the investigation or the justification for not investigating, the reportability decision with its reasoning and date, and any resulting action? Then ask your sales and customer service teams how they log feedback. If any of it bypasses the complaint process, that is a finding waiting to happen.

    Records that sit around these three

    Following a device through these records usually touches others, so have them in order too:

  • Feedback (8.2.1). Production and post-production feedback should feed into risk management.

  • Suppliers (7.4.1). Evaluation, selection, monitoring and re-evaluation, in proportion to risk.

  • Competence (6.2). Training records, and evidence that training worked, for the people handling design changes and complaints.

  • Retention (4.2.5). At least the lifetime of the device as you define it, or as regulations require, and not less than two years from release.
  • Run it as a half-day dry run

    You do not need a full audit to test this. Pick one device family and run three traces: label to medical device file, last design change to design file, three complaints to their closure and reportability decisions. Our post on how to run a pre-audit dry run in half a day sets out the tracer method step by step.

    Have someone other than the record owner do the tracing. ISO 13485 clause 8.2.4 is direct on this for internal audit: auditors shall not audit their own work. If you want to check the audit programme behind all of this, our 10-point check of your internal audit programme applies to ISO 13485 too, with the clause references moved to 8.2.4.

    How TAC helps

    We look at your records the way a certification auditor does, because that is the view that matters on the day. For an ISO 13485 surveillance visit, that usually means:

  • A gap analysis of your system against ISO 13485, using your real device files and records.

  • Planning or reviewing your internal audit programme so it samples the right devices and processes.

  • A mock audit before your surveillance visit, following real devices through your records.

  • Support with nonconformities and corrective action, so findings are closed with evidence.

  • Liaison with your certification body where it helps.
  • We focus on the quality management system and audit readiness. For your device's regulatory route or reporting obligations, your regulatory lead and the MHRA guidance are the right sources.

    Next step

    If your ISO 13485 surveillance visit is coming up and you would like a second pair of eyes on your device files, design changes or complaint records, book a consultation with TAC. We will tell you plainly whether you need a mock audit, help with a few records, or nothing at all.

    ---

    Follow us

  • Instagram: @trainingassuranceconsultancy

  • Facebook: facebook.com/trainingassuranceconsultancy

  • LinkedIn: Training Assurance Consultancy
  • Sources

  • ISO: ISO 13485, Medical devices (current edition ISO 13485:2016). https://www.iso.org/iso-13485-medical-devices.html

  • NQA: ISO 13485:2016 Gap Guide (summaries of 4.2.3, 7.3.10, 8.2.2 and 8.2.3). https://www.nqa.com/medialibraries/NQA/NQA-Media-Library/PDFs/NQA-ISO-13485-Gap-Guide.pdf

  • MHRA: First major overhaul of medical device regulation comes into force across Great Britain (16 June 2025). https://www.gov.uk/government/news/first-major-overhaul-of-medical-device-regulation-comes-into-force-across-great-britain

  • MHRA: Medical devices post-market surveillance requirements, vigilance reporting requirements. https://www.gov.uk/government/publications/medical-devices-post-market-surveillance-requirements/vigilance-reporting-requirements

  • TAC: How to run a pre-audit dry run in half a day. https://trainingassuranceconsultancy.com/blog/how-to-run-a-pre-audit-dry-run-in-half-a-day

  • TAC: Your internal audit programme is due for its own audit, a 10-point check. https://www.trainingassuranceconsultancy.com/blog/internal-audit-programme-10-point-check-iso-9001-14001-45001

Related Topics

ISO 13485Medical devicesSurveillance auditMedical device fileDesign and developmentComplaint handlingAudit preparationMock audit